[Studies of image quality and dose in digital subtraction angiography]
H Heckmann1, K F Kamm, S Vetter
1Abteilung Allgemeine Radiologie, Brüderkrankenhaus Trier.
Methods:
The dose needed for high image quality in l.a. DSA, was determined by investigation of the necessary relationship between dose and image quality. This method is based on measurements of the signal-to-noise ratio and of the contrast-detail detectability. It has been tested on two different systems (Philips integris V3000, Philips Diagnost 97).
Results:
Our measurements prove that an image intensifier entrance dose of 1.1 microGy (image intensifier diameter 38 cm) is sufficient for high image quality in DSA. An increased dose value does not lead to an improvement in image information. For details larger than the pixel size the 512 matrix produces a higher signal-to-noise ratio with a higher contrast-detail detectability, whereas higher spatial resolution results by the use of a 1024 matrix. In rotational angiography, high image quality can be achieved at an image intensifier entrance dose of 1.1 microGy. In contrast to sequences without rotation, image quality is decreased through movement unsharpness. As the maximum exposure time 25 ms should be selected. The use of a 1024 matrix does not lead to a better spatial resolution in rotational angiography.
More Related Videos
05:26Two-Dimensional X-Ray Angiography to Examine Fine Vascular Structure Using a Silicone Rubber Injection Compound
Published on: January 7, 2019
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies V: Intravenous Urography and Retrograde Pyelography
Imaging Studies VII: Vascular Imaging