Related Experiment Videos
[Dosage values for characterization of patient exposure in computerized tomography and their significance for risk
N Hidajat1, R J Schröder, M Wolf
1Strahlenklinik und Poliklinik, Virchow-Klinikum, Medizinischen Fakultät, Humboldt-Universität zu Berlin.
Abstract:
In order to evaluate patients' exposure to radiation in computed tomography, dose quantities such as the computed tomography dose index (CTDI), multiple scan average dose (MSAD) and dose free in air on the axis of rotation are used. The CTDI and the MSAD, derivable from the CTDI, are a good measure for the absorbed doses in the examined volume, but they do not take the biological sensitivity of the organs into account and do not describe the radiogenic risk that is actually relevant for patients and doctors. The dose on the axis of rotation is not an accurate measure of the effective dose and the radiogenic risk. The declaration of a limit for the dose on the axis of rotation should take the different regions into account, in order to guarantee acceptable image quality on one hand and to avoid an unnecessarily high risk on the other side. As physical dose quantities, the CTDI, MSAD and dose on the axis of rotation are useful to characterize and differentiate between programs and systems concerning radiation exposure.