E L Masterson1, B A Masri, C P Duncan
1Department of Orthopaedics, Vancouver Hospital and Health Sciences Centre, British Columbia, Canada.
Abstract:
The postoperative radiographs of 35 patients who underwent impaction allografting of the proximal femur were reviewed. Of Gruen zones that could be clearly visualized, 39.9% contained areas where cement was absent. Even when an adequate mantle was present, cement voids were commonly seen. These cement mantle deficiencies were confirmed in a series of cadaveric impaction allografting procedures. They appear to be a consequence, at least in part, of an inadequate differential between trial and actual component sizes. Additionally, 4 patients were identified with significant component migration secondary to radiographically visible cement mantle fractures within the first 6 months of surgery. It is concluded that the surgical technique requires modification to ensure a more consistent cement mantle and clinical result.