Characteristics of mixed venous oxygen saturation and physical activity as parameters for artificial heart control
T Ohno1, T Masuzawa, M Nakamura
1Department of Artificial Organs, National Cardiovascular Center Research Institute, Osaka, Japan.
Abstract:
Mixed venous oxygen saturation (SvO2) and physical activity (PhyAc) are practical candidates as parameters of total artificial heart (TAH) control because SvO2 can be measured through a transparent blood pump housing with infrared rays and PhyAc can be calculated from signals of an accelerometer used for rate response pacemakers. Although the methods for measurement of the parameters have already been developed, characteristics of these parameters for TAH control, such as during exercise, are still unclear and were examined in this study. SvO2, cardiac output (CO), and PhyAc were measured as parameters. Multi-stage treadmill exercise tests were performed. Difference values (DVs) from the value at the start of exercise showed better correlation than did absolute values. Correlation coefficients between DV in CO and DV in SvO2 and between DV in CO and PhyAc were high at -0.82 and 0.72, while the time constants for the change of SvO2 and CO to the PhyAc change were 26 and 32 sec. Although the correlation coefficient between the CO and SvO2 was higher than that between CO and PhyAc, PhyAc responded more quickly to the speed change compared with the response of SvO2 and CO. It was concluded that SvO2 and PhyAc were useful parameters with different characteristics for TAH control during exercise.
More Related Videos
04:20Integration of Brain Tissue Saturation Monitoring in Cardiopulmonary Exercise Testing in Patients with Heart Failure
Published on: October 1, 2019
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Pathophysiology of Cardiac Performance
Exercise and Cardiac Output
Sustained exercise increases the muscles' oxygen demand, which can be met...