Highly sensitive high-performance liquid chromatographic-tandem mass spectrometric method for the analysis of
T H Eichhold1, M Quijano, W L Seibel
1Procter & Gamble Pharmaceuticals, Mason, OH 45040, USA.
Abstract:
A stable-isotope-dilution HPLC-tandem mass spectrometry-based method was developed for the determination of dextromethorphan in human plasma. Plasma samples were prepared for analysis by solid-phase extraction on octadecylsilane extraction cartridges. Dextromethorphan and the deuterium-labeled dextromethorphan internal standard were chromatographed on a short reversed-phase column and detected by a selected-reaction-monitoring scheme. Linear standard curves were obtained over three orders of magnitude and the limit of quantitation for dextromethorphan was 50 pg/ml, using a 1-ml plasma sample. The combination of HPLC and electrospray tandem mass spectrometry resulted in a rapid, selective and sensitive method for the analysis of dextromethorphan in plasma. The method was applied for the evaluation of the pharmacokinetic profile of dextromethorphan in human volunteers following peroral administration.