Related Experiment Video
Updated: Mar 26, 2026

Disentangling High Strength Copolymer Aramid Fibers to Enable the Determination of Their Mechanical Properties
Published on: September 1, 2018
The Frank Stinchfield Award. Contribution of cable debris generation to accelerated polyethylene wear
J D Hop1, J J Callaghan, J P Olejniczak
1Department of Orthopaedic Surgery, University of Iowa College of Medicine, Iowa City 52242, USA.
Abstract:
The decision of the senior author of a large total hip replacement practice to switch from wire to braided cable for reattachment of the greater trochanter provided the opportunity to evaluate the long term effects (on acetabular component wear, osteolysis, and component loosening) caused by the introduction of metallic debris (generated by cable fretting and breakage) into the total hip arthroplasty construct. Seven hundred and nine consecutive primary total hip arthroplasties were performed during a 5-year period and followed up for a minimum of 10 years. Wire and cable reattachment of the greater trochanter was used sequentially. At minimum 10-year followup the cable group had significantly more wear, osteolysis, and acetabular radiographic evidence of loosening. Those involved in the design and use of total hip arthroplasty devices must minimize potential sources of metallic debris and other potential sources for third body wear in the total hip arthroplasty construct to help ensure longevity of the arthroplasty.
Related Concept Videos
Cable Subjected to Concentrated Loads
Cable Subjected to a Distributed Load
Plastic Deformation in Circular Shafts
Cable Subjected to Its Own Weight
A generalized loading function is employed to analyze a cable subjected to its own weight. This function considers the force acting along the cable's arc length rather than its projected length, providing a more accurate...
Energy Stored In A Coaxial Cable
In the simplest form, a coaxial cable can be represented by two long hollow concentric cylinders in which the current flows in opposite directions. The magnetic field inside and outside the coaxial cable is determined by using Ampère's law. The magnetic field inside...
Fiber Reinforced Concrete