[Possibilities for interpreting digital migration analysis of cement-free PM total hip endoprostheses]
C Hendrich1, C P Rader, G Klein
1Orthopädische Universitätsklinik Würzburg, König-Ludwig-Haus.
Purpose:
Aim of this study was the validation of a digital measurement device for hip implants.
Methods:
From 43 patients with a minimum of four subsequent roentgenograms 246 roentgenograms of PM-shafts (an average 5.72/shaft) and 142 roentgenograms of PM-cups (an average 6.45/cup) were digitized on a DiagnostiX 2048 basis station (Pace Systems, Germany). Modified Sutherland/Nunn methods were chosen for measuring purposes. Regression analysis was done and the results were correlated to implant loosening.
Results:
Concerning the total vertical migration of the cup there is a significant (p < 0.04) difference between the loosened components (6.02 +/- 2.59 mm) and the fixed cups (2.26 +/- 2.89 mm). The average annual vertical migration rate of the stem shows a highly significant difference (p < 0.007) when comparing loosened stems (0.82 +/- 0.43 mm, total migration 4.78 +/- 2.89 mm) and fixed ones (0.20 +/- 0.54 mm, total migration 1.72 +/- 3.02 mm)
Conclusions:
Using the regression analysis and a sufficient number of pictures the system allows us to give migration values that make loosening likely. However for the single implant an individual prognosis of loosening is not possible.