Related Experiment Video
Updated: Jun 29, 2026

09:10
Performing Intracochlear Electrocochleography During Cochlear Implantation
Published on: March 8, 2022
The advanced Combi 40+ cochlear implant
C M Zierhofer1, I J Hochmair, E S Hochmair
1Institute for Applied Physics/Microelectronics, University of Innsbruck, Austria.
The American Journal of Otology
|December 10, 1997
Summary
This study introduces a 12-channel cochlear implant (CI) capable of high-rate pulsatile stimulation. The device enables precise control over stimulation parameters and includes advanced features like self-calibrating telemetry for improved auditory prosthetics.
Area of Science:
- Biomedical Engineering
- Neuroscience
- Audiology
Background:
- Cochlear implants (CIs) are vital for restoring hearing in individuals with severe to profound sensorineural hearing loss.
- Advancements in CI technology focus on enhancing speech perception and sound quality through improved stimulation strategies.
- High-rate pulsatile stimulation offers potential benefits for more natural auditory processing.
Purpose of the Study:
- To present a novel 12-channel cochlear implant (CI) designed for high-rate pulsatile stimulation strategies.
- To detail the technical specifications and capabilities of the developed CI system.
- To introduce integrated features for enhanced performance and safety.
Main Methods:
- Development of a 12-channel CI system capable of generating symmetric biphasic current pulses up to 18.18 kpulses/second.
- Implementation of transcutaneous data and power transfer via a single radiofrequency (RF) channel at 600 kBit/second.
- Integration of a mixed analog/digital CMOS-ASIC for stimulus generation and a self-calibrating back telemetry system.
Main Results:
- The CI system supports stimulation pulse amplitudes from 1.5 microA to 1.5 mA.
- A fully digital data transfer format and a single RF channel facilitate efficient data and power transmission.
- The implant includes safety features like output capacitors and a telemetry system for impedance and field distribution estimation.
- The compact ceramic package measures 33.50 x 23.40 x 3.95 mm3.
Conclusions:
- The presented 12-channel CI is a sophisticated auditory prosthetic device optimized for high-rate pulsatile stimulation.
- The integrated digital data transfer, RF power/data link, and self-calibrating telemetry represent significant technological advancements.
- This CI technology holds promise for improving hearing restoration outcomes and addressing the needs of CI users.