Related Experiment Videos
Speech intelligibility as a function of the number of channels of stimulation for normal-hearing listeners and
1Arizona State University, Department of Speech and Hearing Science, Tempe 85287-0102, USA.
Objective:
One goal was to determine for normal-hearing listeners the number of channels of stimulation necessary to achieve a high level of speech understanding. The second goal was to determine whether patients with a six-channel cochlear implant could achieve the same level of speech understanding as normal-hearing subjects listening to speech processed through six channels.
Methods:
Speech signals were processed, for normal-hearing listeners, either in the manner of cochlear-implant processors with 2-9 fixed channels, or in the manner of a processor which picked, on each update cycle, 6 of 16 channels.
Results:
For the most difficult test material eight fixed channels were necessary to achieve the level of performance achieved with the "n of m" processor. Some cochlear implant patients with a six-channel continuous interleaved sampling processor achieved the same level of performance as normal-hearing subjects listening to speech via six channels.
Conclusions:
A signal processor for cochlear implants with eight channels should produce the same level of intelligibility as a processor with many more channels. Processors using continuous interleaved sampling technology can provide a signal which results in the same level of speech understanding as normal, acoustic stimulation.