Related Experiment Videos
The influence of bone quality on Periotest values of endosseous dental implants at stage II surgery
R S Truhlar1, F Lauciello, H F Morris
1Department of Veterans Affairs Medical Center, Northport, NY, USA.
Abstract:
Periotest values (Periotest, Siemens AG, Bensheim, Germany) were recorded as a baseline variable at surgical uncovering in the ongoing multicenter, prospective clinical studies of the Dental Implant Clinical Research Group, which uses implants from the Spectra-System (Core-Vent Corporation, Las Vegas, NV). For 2,212 osseointegrated implants, the mean Periotest value (PTV) of mandibular implants was -4.14 (anterior, -4.22; posterior, -4.06) versus -3.24 for maxillary implants (anterior, -2.91; posterior, -3.91). Implants in the densest bone (quality 1) had the lowest mean PTV (-4.13), followed by quality 2 (-4.00), quality 3 (-3.58), and quality 4 (-2.64).