Uncemented total hip replacements using the Mecron acetabular cup--a prospective study

H H Lim1, D M Hunt

  • 1Department of Orthopaedics, St Mary's Hospital, London, United Kingdom.

Annals of the Academy of Medicine, Singapore
|July 1, 1997
PubMed
Summary

The Mecron acetabular threaded cup showed poor short-term results in primary total hip replacements, with 5 revisions due to loosening. Migration and tilting led to the prosthesis being abandoned.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

A pre-admission triaging tool to predict severe COVID-19 cases: ABCD score.

The Medical journal of Malaysia·2022
Same author

Visual system evolution and the nature of the ancestral snake.

Journal of evolutionary biology·2015
Same author

Descending and tonotopic projection patterns from the auditory cortex to the inferior colliculus.

Neuroscience·2015
Same author

Association of fatty acid composition in serum phospholipids with metabolic syndrome and arterial stiffness.

Nutrition, metabolism, and cardiovascular diseases : NMCD·2011
Same author

Identification and characterization of visual pigments in caecilians (Amphibia: Gymnophiona), an order of limbless vertebrates with rudimentary eyes.

The Journal of experimental biology·2010
Same author

Optoacoustic induced vibrations within the inner ear.

Optics express·2010

Area of Science:

  • Orthopedic surgery
  • Biomaterials science
  • Medical device engineering

Background:

  • Uncemented threaded cups have gained popularity for hip replacements.
  • The Mecron acetabular threaded cup system was evaluated for primary hip arthroplasty.

Purpose of the Study:

  • To assess the short-term outcomes of the Mecron acetabular threaded cup system in primary total hip replacements.
  • To identify failure modes and complications associated with this specific implant.

Main Methods:

  • A series of 104 primary total hip replacements using the Mecron acetabular cup and Ring system femoral stem were performed between 1987 and 1991.
  • Patients were followed annually for up to 4 years, with assessments including the modified Harris score and radiographic analysis.

Related Experiment Videos

Main Results:

  • Five hips (4.8%) required revision surgery, all due to acetabular loosening.
  • 60.6% of cases had excellent or good short-term results, while 18.3% had poor results.
  • Radiographic evaluation revealed migration and tilting of the acetabular cup in all cases; heterotopic bone formation occurred in 10 patients.

Conclusions:

  • The Mecron acetabular cup demonstrated significant issues with loosening, migration, and tilting, leading to its discontinuation.
  • These findings highlight the importance of implant design and fixation in achieving long-term success in total hip arthroplasty.