The response of the Intoxilyzer 5000 to five potential interfering substances
1Department of Chemistry, University of Waikato, Hamilton, New Zealand.
Abstract:
A study was conducted of potential vapor phase interferents which could be present on human breath and also be capable of inducing a false-positive response for ethanol on the evidential infrared-based breath testing device, the Intoxilyzer-5000. This involved preparation and validation of a range of vapor standards, which were introduced to the instrument using a dynamic flow double-bubbler system. Potential interferents were chosen on the basis of both their infrared signatures and their general availability, and included toluene, m-xylene, o-xylene, methanol and isopropanol. All compounds tested were found to be capable of inducing false-positive readings for ethanol in a highly reproducible manner, as a result of which it has been possible to derive precise least-squares equations describing the ethanol readout expected for any given blood concentration of toluene, m-xylene, o-xylene, methanol and isopropanol. The likelihood of an interference compromising the integrity of the analysis is related to both the toxicological significance and prevalence of a given blood concentration of each solvent, and the point at which the instrumental interference light is triggered in each case.
More Related Videos
11:38High Content Screening Analysis to Evaluate the Toxicological Effects of Harmful and Potentially Harmful Constituents (HPHC)
Published on: May 10, 2016
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Gas Chromatography: Types of Detectors-II
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Atomic Emission Spectroscopy: Interference