Anaesthetic pollution. Potential sources, their identification and control

J P Barker1, M O Abdelatti

  • 1Edgware General Hospital, Middlesex, UK.

Anaesthesia
|December 24, 1997
PubMed
Summary

Anaesthetic gas pollution in operating rooms stems from various sources beyond the breathing system. Identifying and controlling these leaks using infra-red analyzers is crucial for protecting healthcare worker health.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Gas exchange parameters in radiotherapy patients during breathing of 2%, 3.5% and 5% carbogen gas mixtures.

The British journal of radiology·2001
Same author

A cuff pressure controller for tracheal tubes and laryngeal mask airways.

Anaesthesia·1999
Same author

Modification of a simple algometer.

Anaesthesia·1998
Same author

Ondansetron compared with metoclopramide in the treatment of PONV.

British journal of anaesthesia·1998
Same author

Management of hypoxaemia during one-lung anaesthesia--a suggestion.

Anaesthesia·1998
Same author

A cuff inflator for tracheal tubes.

Anaesthesia·1997

Area of Science:

  • Occupational Health
  • Environmental Health
  • Anesthesiology

Background:

  • Anaesthetic gases can escape into the operating room environment from multiple sources, not just the anaesthetic breathing system.
  • Regulations like COSHH and OES mandate risk assessment and control of anaesthetic exposure for healthcare workers.
  • Scavenging equipment, while installed, may not be fully effective in controlling anaesthetic pollution.

Purpose of the Study:

  • To identify and assess controllable sources of anaesthetic gas pollution in an operating theatre.
  • To evaluate the effectiveness of current pollution control measures.
  • To recommend methods for improving the control of anaesthetic pollution.

Main Methods:

  • Utilized an infra-red analyser for leak testing and background monitoring during an initial COSHH assessment.

Related Experiment Videos

  • Investigated various potential pollution sources including vaporiser filling, induction techniques, face mask leaks, and equipment integrity.
  • Monitored specific areas such as gas outlets and connections within the anaesthetic system.
  • Main Results:

    • Identified several controllable sources of anaesthetic gas pollution, including ill-fitting face masks, loose breathing system connections, and laryngeal mask leaks.
    • Detected a significant leak of nitrous oxide (N2O) near a multigas monitor's unscavenged gas outlet.
    • Highlighted that pollution sources remote from scavenging equipment can contribute to ineffective pollution control.

    Conclusions:

    • Infra-red leak testing is an effective method for identifying hidden sources of anaesthetic gas pollution.
    • Many identified pollution sources are amenable to control and remedy, improving occupational safety.
    • Recommendations include implementing regular leak testing to enhance the management of anaesthetic gas exposure.