Ultrasound-mediated destruction of contrast agents. Effect of ultrasound intensity, exposure, and frequency
K W Walker1, G A Pantely, D J Sahn
1Division of Cardiology, Medical University of South Carolina, Charleston 29425-2221, USA. walkerkw@musc.edu
Rationale And Objectives:
Although ultrasound contrast microbubbles theoretically could serve as tracers for the noninvasive quantification of blood flow, results have been inconsistent. Accurate quantification may be limited by ultrasound energy-mediated microbubble destruction. This study examined the effect of different ultrasound delivery parameters on microbubble destruction.
Methods:
Experiments were performed in an in vitro hydraulic perfusion model consisting of a thin-walled rubber tube encased in agar. Ultrasonic parameters tested during different parts of the experiment were (1) intensity, (2) duration, and (3) frequency. Four ultrasound contrast agents: Aerosomes MRX115 (ImaRx Pharmaceuticals Corp., Tucson, AZ), Imagent AF0150 US (Alliance Pharmaceutical Corp., San Diego, CA), Levovist (Berlex Laboratories, Wayne, NJ), and Echogen (Sonus Pharmaceuticals, Bothel, WA) were imaged with three different ultrasound systems: ATL Ultramark AM-9 HDI, Vingmed 800 and Hewlett-Packard 2500.
Results:
Microbubble destruction and reductions in reflectivity were noted in all agents tested. Although no significant reductions in counts or reflectivity occurred at 0.3 W/cm2 with any agent, exposure to 25 W/cm2 produced more than 80% reductions in both microbubble counts (P < 0.0001) and reflectivity (P < 0.0001). Declines in reflectivity were increased by longer exposure to ultrasound (P < 0.0001); slower flow through an ultrasound beam (P < 0.0001); continuous, rather than intermittent, imaging (P = 0.0002); use of a higher pulse repetition rate (P < 0.0001); and exposure to 2.5 MHz, rather than 7.5 MHz, ultrasound (P < 0.0001).
Conclusions:
Ultrasound energy-mediated destruction of contrast microbubbles is a function of many factors, including ultrasound intensity, duration, and frequency. Optimization of ultrasound delivery parameters may be used to maximize or minimize the destruction of ultrasound contrast agents.
More Related Videos
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Imaging Studies III: Computed Tomography