Noncorneal astigmatism related to poly(methyl methacrylate) and plate-haptic silicone intraocular lenses
Purpose:
To evaluate noncorneal astigmatism after implantation of a one-piece, plate-haptic silicone or one-piece poly(methyl methacrylate) PMMA intraocular lens (IOL).
Setting:
Klinik und Poliklinik für Augenheilkunde der Universität Regensburg, Germany.
Methods:
After a follow-up of at least 3 months, the degree of postoperative noncorneal astigmatism was calculated using the data from an auto refractometer. Sixty patients were evaluated: 30 with silicone IOLs (Group A) and 30 with PMMA lenses (Group B).
Results:
Mean noncorneal astigmatism in Group A was 0.78 diopter (D) +/- 0.51 (SD), which was statistically significantly higher than that in Group B (0.51 +/- 0.27 D) (P = .013). The highest noncorneal astigmatism, 2.32 D, occurred in Group A.
Conclusions:
Based on these findings, we recommend noncorneal astigmatism be considered in eyes with a one-piece, plate-haptic silicone IOL and postoperative astigmatism. A prospective study with a standardized capsulorhexis size is needed to ascertain whether this IOL-related noncorneal astigmatism is caused by capsular shrinkage, which may then be treatable with a laser capsulotomy of the anterior capsule.