Related Experiment Video
Updated: Aug 16, 2026

Chemically-blocked Antibody Microarray for Multiplexed High-throughput Profiling of Specific Protein Glycosylation in Complex Samples
Published on: May 4, 2012
Glycohemoglobin filter assay for doctors' offices based on boronic acid affinity principle
F Frantzen1, K Grimsrud, D E Heggli
1Axis Nord, Mørkved, Norway. frank.frantzin@axisbio.no
Abstract:
We present a new filter assay for the determination of glycohemoglobin as a unique application of the boronic acid affinity principle. With the use of a water-soluble blue-colored boronic acid derivative and a specific precipitation method for hemoglobin, total hemoglobin including bound boronic acid is precipitated and collected on a filter strip before quantification. Hemoglobin and boronic acid are quantified by a dual-wave-length reflectometric measurement, and the result is reported directly as percent glycohemoglobin. The test is simple, quick, and designed as a doctors' office test for the monitoring and management of diabetes. The imprecision of the assay is < 4% over the range 3-18% Hb A1c, and the method is linear up to at least 20% Hb A1c. Comparisons with four well-established glycohemoglobin methods yielded correlation coefficients ranging from 0.94 to 0.99, with slopes from 0.94 to 1.01.