Related Experiment Video
Updated: Jul 25, 2026

Optical Detection of E. coli Bacteria by Mesoporous Silicon Biosensors
Published on: November 20, 2013
Rapid detection of D-dimer using a fiber optic biosensor
C A Rowe1, J S Bolitho, A Jane
1Naval Research Laboratory, Center for Bio/Molecular Science and Engineering, Washington, DC 20375-5438, USA.
Abstract:
We describe a rapid and sensitive method for detection and quantification of D-dimer and other crosslinked fibrin degradation products (XL-FDPs), which are present in elevated concentrations in patients with sepsis and thrombotic disorders. The method utilizes a sandwich fluoroimmunoassay immobilized in the sensing region of an evanescent wave biosensor. Physiological concentrations of D-dimer and high molecular weight XL-FDP could be determined in buffer and plasma samples on calibrated fibers in 11 min. Samples from septic patients were assayed using ELISA and the fiber optic method; concentrations determined by fiber optic assay were strongly correlated with those determined by ELISA (r = 0.918); intra- and inter-assay errors were comparable to those from ELISAs. Given its accuracy and rapid response time, this fiber optic biosensor shows great potential for development as a diagnostic tool.