This study explored hemofiltration efficacy using a novel balancer and peritoneal dialysis machine for substitute delivery. The methods proved effective for managing fluid balance and electrolyte disorders in patients undergoing hemofiltration.
Area of Science:
- Nephrology
- Biomedical Engineering
- Renal Replacement Therapy
Context:
- Hemofiltration is a crucial renal replacement therapy.
- Managing fluid balance and electrolyte disorders is essential in patients with kidney disease.
- Existing hemofiltration protocols require optimization for efficiency and cost-effectiveness.
Purpose:
- To evaluate the antiuremic efficacy of hemofiltration using a self-made balancer.
- To assess the utility of an APD-02 peritoneal dialysis machine for substitute delivery in hemofiltration.
- To investigate the feasibility of F-60 hemofiltration cartridge reuse.
Summary:
- 47 hemofiltration procedures were conducted on 10 patients.
- A custom-designed balancer, leveraging volumic hydrodynamics and filtrate-dependent substitution, was employed.
- The APD-02 peritoneal dialysis machine provided bicarbonate substitute, aiding in the prophylaxis of osmodisturbances and dehydration.
Impact:
- The study demonstrated the successful application of a novel hemofiltration technique.
- The developed technology for F-60 cartridge reuse was validated, suggesting potential cost savings.
- One patient underwent an extended isolated hemofiltration program for over two months, highlighting the therapy's potential for long-term management.