Simplified implantation of single-lead pectoral cardioverter defibrillators using device-based testing
1Department of Cardiology, Hospital of the Philipps-University of Marburg, Germany.
The American Journal of Cardiology
|March 4, 1998
Abstract:
Device-based testing of single-lead pectoral defibrillators (defibrillation efficacy testing without an external defibrillation system after complete implantation of the device) resulted in an adequate defibrillation threshold (< or = 20 J) in 45 of 50 study patients (90%). Mean surgical implantation time (skin to skin) was 62 +/- 29 minutes without perioperative mortality and without implantable cardioverter defibrillator infection during follow-up. Thus, device-based testing appears to be a simple and safe method to test defibrillation efficacy of single-lead pectoral defibrillators.