Spetzler titanium aneurysm clips: compatibility at MR imaging
Related Experiment Video
Updated: Oct 30, 2026

Evaluation of a Novel Laser-assisted Coronary Anastomotic Connector - the Trinity Clip - in a Porcine Off-pump Bypass Model
Published on: November 24, 2014
1Department of Radiology, School of Medicine, University of Southern California, Los Angeles, USA.
Abstract:
To evaluate the compatibility of a commercially pure titanium aneurysm clip associated with magnetic resonance (MR) imaging at 1.5 T. Artifacts of the Spetzler titanium aneurysm clip were compared with those produced by six different nonferromagnetic aneurysm clips. With the titanium clip, no magnetic attraction was present, heating was minor, and the artifacts involved a small signal void. With the six other aneurysm clips, artifacts were larger. The presence of Spetzler titanium aneurysm clips is safe at MR imaging at 1.5 T or less.