Digital cellular telephone interaction with implantable cardioverter-defibrillators
J G Fetter1, V Ivans, D G Benditt
1Medtronic, Inc, Minneapolis, Minnesota 55432, USA.
Journal of the American College of Cardiology
|March 21, 1998
Summary
Electromagnetic interference from digital cellular phones did not affect implantable cardioverter-defibrillators (ICDs) in clinical tests. However, proximity to the device may activate a reed switch, temporarily suspending detection.
Area of Science:
- Biomedical Engineering
- Cardiology
- Electromagnetics
Background:
- Implantable cardioverter-defibrillators (ICDs) are crucial for managing life-threatening arrhythmias.
- Mobile cellular telephones generate electromagnetic interference (EMI) that could potentially affect electronic medical devices.
- Understanding potential interactions between cellular phones and ICDs is vital for patient safety.
Purpose of the Study:
- To assess in vivo and in vitro whether EMI from North American Digital Communications (NADC)/Time Division Multiple Access-50-Hz (TDMA-50) mobile cellular telephones interferes with ICD operation.
- To verify these findings using a selection of telephones representing worldwide systems.
Main Methods:
- In vivo testing involved 41 patients with Medtronic ICDs exposed to an AT&T 6650 digital telephone (NADC/TDMA-50).
- In vitro testing included antenna-generated far-field exposure and near-field exposure using analog and digital cellular telephones.
- Testing evaluated ICD response to EMI under various simulated conditions.
Main Results:
- No oversensing or interference with ICD operation was observed in any of the 41 patients during the in vivo study.
- In vitro far-field testing showed no ICD sensing interference from TDMA-50 and Global System Mobile (GSM) telephone modulation frequencies.
- Near-field tests indicated no interaction with normal ICD operation, except for static magnetic fields activating the reed switch at close proximity (< or = 0.5 cm).
Conclusions:
- TDMA-50 cellular telephones did not demonstrate interference with the tested ICD models.
- A recommendation is made to maintain a distance of at least 15 cm (6 inches) between digital cellular telephones and ICDs.
- While EMI is not an issue, the static magnetic field's effect on the reed switch warrants caution regarding device placement.