Proposal for standardizing direct-reading biosensors for blood glucose
1Department of Clinical Biochemistry, Herlev Hospital, Denmark. nfa@post6.tele.dk
Clinical Chemistry
|March 25, 1998
Summary
Direct-reading glucose biosensors measure molality. Converting results using water content ratios aligns direct-reading biosensor measurements with routine methods for blood glucose concentration. Standardization to plasma-equivalent values is recommended.
Area of Science:
- Clinical Chemistry
- Biosensor Technology
- Analytical Chemistry
Background:
- Direct-reading glucose biosensors measure molality, differing from concentration-based methods.
- Aqueous calibration can lead to higher glucose readings in blood and plasma compared to routine methods.
- This discrepancy is theoretically explained by the ratio of water concentrations between calibrator and sample.
Purpose of the Study:
- To confirm the theoretical explanation for higher glucose readings from direct-reading biosensors.
- To evaluate the accuracy of direct-reading glucose biosensor measurements against routine methods.
- To propose a method for standardizing glucose reporting to avoid confusion.
Main Methods:
- Glucose levels were measured in 14 blood and 40 plasma samples using the Chiron Model 860 direct-reading glucose sensor.
- Measurements were compared with results from a routine method (ESAT 6660; Eppendorf).
- Chiron sensor results were converted using assumed water concentrations for blood (0.84 kg/L) and plasma (0.93 kg/L) relative to the calibrator's water concentration (0.99 kg/L).
Main Results:
- Conversion of direct-reading glucose biosensor results using water content ratios achieved agreement with the routine method.
- Individual conversion based on hematocrit in whole blood samples was less accurate.
- The study confirmed that molality measurement by direct-reading biosensors leads to higher results than concentration-based methods.
Conclusions:
- Standardizing the reporting of whole-blood glucose results as equivalent plasma concentrations is proposed to reduce confusion.
- This standardization can be conveniently achieved using a reference material like NIST SRM 965.
- Accurate glucose measurement and reporting are crucial for clinical decision-making.