A field method for blink reflex latency R-1 (BRL R-1) and prediction equations for adults and children
K H Kilburn1, J C Thornton, B Hanscom
1Environmental Sciences Laboratory, University of Southern California School of Medicine, Los Angeles, USA.
Purpose:
To develop a field worthy apparatus for blink reflex latency R-1 (BRL R-1), to compare mechanical and electrical stimulation, to define a standardized measurement of BRL R-1, to test a field worthy procedure and to produce population-based prediction equations.
Methods:
A special low voltage electromyographic (EMG) recording system was constructed which is uninfluenced by electrical and magnetic field. The supraorbital notches were stimulated mechanically using a small hammer and electrically using a stimulus of 2.0 ms duration and 2-5 mA 13. A computer program was developed which substracts the baseline for all data points and sums R-1 for 10 tests by identifying the noise floor or baseline of the EMG recording and the noise envelope. The beginning of R-1 and R-2 is defined as 2 times the noise amplitude of this envelope. BRL R-1 was compared after tap and after electrical stimulation in 16 subjects. Prediction equations were developed from data of 240 adults and 163 children unexposed to chemicals.
Results:
Mechanically and electrically elicited mean BRL R-1 in 16 subjects was identical at 12.6 ms on the right and 12.7 ms vs 12.5 on the left. In adults, BRL R-1 depended on age. In children BRL R-1 was age, weight and height dependent.
Conclusion:
BRL R-1 elicited by supraorbital tap is field worthy, computer defined and identical to electrical BRL.