Carbamazepine interference with an immune assay for tricyclic antidepressants in plasma
D S Chattergoon1, Z Verjee, M Anderson
1Hospital for Sick Children, University of Toronto, Ontario, Canada.
Background:
Drug toxicological screening is commonly used as a diagnostic tool in patients with suspected toxic ingestion. False positive results due to cross-reactive compounds in drug assays may lead to misdiagnosis and mismanagement, especially when child abuse is suspected.
Case Report:
Two of our patients with history of ingestion of carbamazepine were tested positive on screening with the tricyclic antidepressant immunoassay. The immunoassay's known cross-reactivity for carbamazepine is reportedly as low as 0.3%. Plasma samples of our patients were initially considered positive for tricyclic antidepressants because the cross-reaction of carbamazepine gave tricyclic antidepressant concentrations as imipramine equivalent sufficiently above the assay cut-off point (20 ng/mL). Later, confirmatory urine testing of both patients using high-performance liquid chromatography was negative for tricyclic antidepressants.
Conclusion:
This interference has significant clinical implications, and can be avoided on urine testing using a specific chromatographic assay such as high-performance liquid chromatography.
More Related Videos
12:21Thermostabilization, Expression, Purification, and Crystallization of the Human Serotonin Transporter Bound to S-citalopram
Published on: November 27, 2016
06:50A Soluble Tetrazolium-Based Reduction Assay to Evaluate the Effect of Antibodies on Candida tropicalis Biofilms
Published on: September 16, 2022
Related Concept Videos
Drugs Affecting Neurotransmitter Release or Uptake
Antidepressant Drugs: Tricyclics, SSRIs, and SNRIs
Therapeutic Drug Monitoring: Affecting Factors