Dose from syringe procedures during technetium-99m radiopharmaceutical preparation
S G Prussin1, G Theofanous, D Casey
1Department of Nuclear Engineering, University of California Berkeley 94720, USA.
Unlabelled:
We examined the dose from 99mTc contained in syringes and shielded vials to assess in detail the dose burdens in a central pharmacy.
Methods:
Absorbed-dose rates at the end of the plunger of a shielded syringe, when 99mTc is contained either in the syringe or in a shielded vial from which the activity would be drawn, were measured with CaF2 dosimeters. The dose rates also were calculated with a Monte Carlo model.
Results:
When activity was contained in either 3-ml or 10-ml disposable syringes shielded by lead glass, the absorbed-dose rates were 1.35-1.62 mGy hr-1 GBq-1 (5-6 mrad hr-1 mCi-1). When the activity was contained in either a shielded elution or product kit vial, the absorbed-dose rates at the end of the syringe plunger were about 0.40 mGy hr-1 GBq-1 (1.4-1.5 mrad hr-1 mCi-1). These results were reproduced with reasonable accuracy by Monte Carlo simulations.
Conclusion:
The dose burden per unit of activity handled from 99mTc in procedures using syringes is likely to be two to five times larger than the dose burden from calibrating generator eluate. The Monte Carlo simulations suggest that lead K x-rays may be responsible for a significant fraction of the total dose to the fingers and hand of the pharmacist when lead-glass syringe shields are used.
More Related Videos
09:03Radiotracer Administration for High Temporal Resolution Positron Emission Tomography of the Human Brain: Application to FDG-fPET
Published on: October 22, 2019
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Related Concept Videos
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography