Rapid, automated assay for progesterone on the Abbott AxSYM analyzer
D H Wilson1, W Groskopf, S Hsu
1Department of Fertility, Pregnancy, and Neurodiagnostics, AxSYM R&D, Abbott Laboratories, Abbott Park, IL 60064, USA. wilson@apmac.abbott.com
Clinical Chemistry
|April 29, 1998
Summary
This study introduces an automated progesterone (P4) assay on the Abbott AxSYM analyzer, offering high throughput and accuracy for clinical diagnostics. The assay demonstrates excellent precision, minimal interference, and strong correlation with existing methods.
Area of Science:
- Clinical Chemistry
- Immunoassay Development
- Biomarker Analysis
Background:
- Progesterone (P4) is a critical hormone in reproductive health and pregnancy.
- Accurate and efficient measurement of P4 is essential for clinical diagnosis.
- Existing methods for P4 analysis may have limitations in throughput or automation.
Purpose of the Study:
- To develop and validate an automated, one-step competitive immunoassay for progesterone (P4) determination.
- To evaluate the performance characteristics of the new assay using the Abbott AxSYM analyzer.
- To assess the assay's suitability for routine clinical laboratory use.
Main Methods:
- A one-step competitive immunoassay format was employed.
- Progesterone (P4) was immobilized on latex microparticles.
- Alkaline phosphatase (AP)-conjugated anti-P4 antibody was used as the detection system.
- Assay performance was evaluated using NCCLS EP5-T protocols and multisite correlation studies.
Main Results:
- The assay demonstrated high precision with total CVs ranging from 3.4% to 8.2%.
- The detection limit was 0.10 microg/L, with excellent recovery (83-116%) and accuracy (92-115%).
- Minimal cross-reactivity (<6.3%) and interference from common biological substances were observed.
- Strong correlation (r=0.976) was found with a commercial RIA method.
- High throughput achieved: >80 tests/hour (batch) and 60 tests/hour (mixed load).
Conclusions:
- The automated Abbott AxSYM P4 assay is precise, accurate, and robust.
- The assay shows minimal interference and cross-reactivity, ensuring reliable results.
- Its high throughput and strong correlation with RIA make it suitable for routine clinical use.