[Comparison between steam and gas sterilized optics. Costs, availability, maintenance]
Summary
Steam sterilization is a safe, fast, and cost-effective method for optical instruments, showing fewer repairs and lower overall costs compared to ethylene oxide sterilization.
Area of Science:
- Medical device sterilization
- Surgical instrument reprocessing
Background:
- Optical instruments are critical for various surgical procedures.
- Sterilization methods impact instrument longevity and hospital costs.
Purpose of the Study:
- To compare the practical performance and cost-effectiveness of ethylene oxide (EO) versus steam sterilization for optical surgical instruments.
- To evaluate criteria including frequency of use, availability, sterilization cycles, and repair data.
Main Methods:
- A comparative study was conducted in a 300-bed acute-care hospital.
- 33 ethylene oxide (EO)-sterilized and 14 steam-sterilized optical instruments were analyzed.
- Data collected included usage frequency, repair history, and associated costs.
Main Results:
- Steam-sterilized instruments (14) were used 1101 times with 8 repairs costing 2000 SFR.
- EO-sterilized instruments (33) were used 1043 times with 10 repairs costing up to 2000 SFR.
- Total costs: 77,000 SFR for steam vs. 132,000 SFR for EO-sterilized instruments.
Conclusions:
- Steam sterilization is a preferred method for optical instruments due to its safety, speed, and lower toxicity.
- Steam sterilization offers advantages in personal care and pollution control.
- The study supports steam sterilization as a reasonable and efficient practice.