The Platelet Function Analyzer (PFA-100): a novel in-vitro system for evaluation of primary haemostasis in children

M D Carcao1, V S Blanchette, J A Dean

  • 1Division of Haematology/Oncology, The Hospital for Sick Children, Toronto, ON, Canada.

Insights

The PFA-100 system measures platelet function via closure time (CT). Neonates show shorter CTs than children and adults, suggesting potential for diagnosing von Willebrand disease in pediatric patients.

Area of Science:

  • Hematology
  • Pediatric Medicine
  • Diagnostic Assays

Background:

  • Primary hemostasis assessment is crucial for diagnosing bleeding disorders.
  • The PFA-100 system offers an in-vitro method to evaluate platelet function.
  • Establishing age-specific reference ranges is essential for accurate interpretation of diagnostic tests.

Purpose of the Study:

  • To establish normal closure time (CT) ranges using the PFA-100 system in healthy children, adults, and neonates.
  • To assess the PFA-100 system's utility in pediatric populations, including those with hemophilia.
  • To evaluate the potential of the PFA-100 system for screening pediatric von Willebrand disease.

Main Methods:

  • Utilized the PFA-100 system for in-vitro assessment of platelet function.
  • Measured closure time (CT) in healthy neonates, children, and adults.
  • Analyzed CTs in relation to age and needle gauge (21G, 23G) for blood collection.
  • Included children with hemophilia in the study cohort.

Main Results:

  • Mean CTs in healthy children were similar to adults and independent of needle gauge.
  • Mean CTs in healthy neonates were significantly shorter than in children and adults.
  • Children with hemophilia exhibited normal CTs on the PFA-100 system.
  • The PFA-100 system demonstrated potential utility in screening for von Willebrand disease in children.

Conclusions:

  • Normal PFA-100 CT ranges vary with age, with neonates having distinct values.
  • The PFA-100 system may not be sensitive for detecting hemophilia in children.
  • The PFA-100 system shows promise as a screening tool for von Willebrand disease in pediatric patients.