Chromogenic assays for equine coagulation factors VII, VIII:C, IX, and X, and C1-esterase inhibitor

M J Topper1, K W Prasse

  • 1Department of Pathology, College of Veterinary Medicine, University of Georgia, Athens 30602, USA.

Insights

Automated human chromogenic assays successfully measure coagulation factors VII, VIII:C, IX, X, and C1-esterase inhibitor in horses. This method provides a reliable tool for evaluating equine hemostasis disorders.

Area of Science:

  • Veterinary Hematology
  • Clinical Pathology
  • Coagulation Assays

Background:

  • Manual chromogenic assays for specific coagulation factors are established in human medicine.
  • Adapting these assays for equine plasma requires validation and automation for routine use.
  • Accurate measurement of coagulation factors is crucial for diagnosing and managing bleeding disorders in horses.

Purpose of the Study:

  • To adapt manual human chromogenic assays for coagulation factors VII (F.VII), VIII:coagulant (F.VIII:C), IX (F.IX), and X (F.X), and C1-esterase inhibitor (C1-INH) for automated analysis.
  • To establish reference ranges for these coagulation factors in healthy horses.
  • To assess the utility of these automated assays for evaluating equine hemostasis.

Main Methods:

  • Human chromogenic assay kits for F.VII, F.VIII:C, F.IX, F.X, and C1-INH were adapted for a computer-assisted automated analyzer.
  • Standards were prepared using pooled citrated equine plasma for F.VII, F.VIII:C, and F.X assays.
  • Standards for F.IX and C1-INH assays utilized commercial pooled citrated human plasma.
  • Assays were performed on plasma from 10 healthy horses to determine ranges and assess reproducibility.

Main Results:

  • Mean activities in equine plasma were: F.VII (226 ± 19%), F.VIII:C (209 ± 31%), F.IX (149 ± 38%), F.X (88 ± 12%), and C1-INH (18.4 ± 8.4%).
  • Intra-assay coefficients of variation (CV) ranged from 2.1% to 5.3%, while inter-assay CVs ranged from 3.8% to 22.0%.
  • The automated assays demonstrated acceptable precision for routine clinical use.

Conclusions:

  • Human chromogenic assay kits can be effectively automated for measuring F.VII, F.VIII:C, F.IX, F.X, and C1-INH activities in equine plasma.
  • These automated assays provide a reliable method for routine assessment of coagulation factor activities in horses.
  • The adapted assays are valuable for the evaluation of horses with hemostatic disorders.