Levels of microbial contamination on surgical instruments
W A Rutala1, M F Gergen, J F Jones
1Division of Infectious Diseases, University of North Carolina School of Medicine, Chapel Hill, USA.
Objective:
To ascertain the microbial load and type of organisms on used surgical instruments following standard cleaning, which consisted of the use of a washer sterilizer followed by sonic cleaning.
Design:
In this prospective experimental study, used surgical instruments were immersed in Peptamin Tween broth, the broth agitated, and then filtered through a 0.45 microm filter. Quantitative cultures were performed, and all microbes were identified by using standard techniques.
Setting:
This study was conducted at a 660-bed university hospital.
Results:
The microbial load remaining on used surgical instruments after cleaning was as follows: 36 (72%) instruments 0 to 10 colony-forming units (CFU), 7 (14%) instruments 11 to 100 CFU, and 7 (14%) instruments > 100 CFU. Organisms contaminating the instruments included coagulase-negative staphylococcus (56%) followed by Bacillus (22%) and diphtheroids (14%). No other microbes were isolated from more than 4% of the instruments.
Conclusion:
Most used nonlumen surgical instruments contain less than 100 CFU of relatively nonpathogenic microorganisms after cleaning. This suggests that new low-temperature sterilization technologies are likely to be highly effective in preventing cross-transmission of infection via nonlumen medical instruments.
Insights
Standard cleaning effectively reduces microbial load on surgical instruments to less than 100 colony-forming units (CFU). This indicates that new sterilization methods will likely prevent cross-transmission of infections from nonlumen instruments.
Area of Science:
- Microbiology
- Infection Control
- Surgical Instrument Reprocessing
Background:
- Surgical instruments require thorough cleaning to prevent healthcare-associated infections.
- Standard cleaning protocols involve washer sterilizers and sonic cleaning.
Purpose of the Study:
- To determine the microbial load and types of microorganisms on used surgical instruments after standard cleaning.
- To assess the effectiveness of washer sterilizer and sonic cleaning methods.
Main Methods:
- Prospective experimental study at a university hospital.
- Used surgical instruments were cultured using Peptamin Tween broth and filtered.
- Quantitative cultures and standard microbiological identification techniques were employed.
Main Results:
- 72% of instruments had 0-10 colony-forming units (CFU) post-cleaning.
- 14% had 11-100 CFU, and 14% had over 100 CFU.
- Coagulase-negative staphylococcus (56%) was the most common organism, followed by Bacillus (22%) and diphtheroids (14%).
Conclusions:
- Most nonlumen surgical instruments harbor fewer than 100 CFU of relatively nonpathogenic microbes after cleaning.
- Current cleaning methods are effective in reducing microbial burden.
- New low-temperature sterilization technologies are expected to be highly effective for nonlumen instruments.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Asepsis
Key Techniques in Microbiology
Methods for Controlling Microbial Growth
Sources of Food Contamination