Cardiac arrhythmias with external fixed-rate signal generators in shock wave lithotripsy with the Medstone
1University of North Carolina, Chapel Hill 27599-7235, USA.
Objectives:
To evaluate the incidence and outcome of cardiac arrhythmias associated with an external fixed-rate generator and the Medstone lithotripter with regard to stone location, anesthesia type, patient age, and gender.
Methods:
Data were obtained on 933 patients who underwent shock wave lithotripsy (SWL) with the Medstone lithotripter. Patients were randomly divided into two study groups, including those with pre-existing cardiac disease, hypertension, and cardiac medications. Group 1 (576 patients) were treated with the external fixed-rate signal generator (cardiac simulator). Group 2 (357 patients) underwent electrocardiogram-gated SWL (no simulator). All patients were monitored for arrhythmias during SWL. Patients who developed arrhythmias on the simulator were converted to gated SWL.
Results:
In group 1, 463 patients (80%) had no arrhythmias during treatment, whereas 113 patients (20%) developed cardiac arrhythmias (primarily premature ventricular contractions [PVCs]) and were successfully converted to gated SWL. Only 1 patient in group 2 (0.3%) had a documented arrhythmia (PVC). Group 1 patients who developed arrhythmias with a cardiac simulator were more likely to have renal rather than ureteral stones. Chi-square analysis showed a significant correlation between patients with and without arrhythmias in group 1 with regard to stone location (P <0.0001). In group 1, regional anesthesia, intravenous anesthesia, and general anesthesia had an arrhythmia rate of 14%, 27%, and 38%, respectively. The only patient in group 2 who developed arrhythmias was under general anesthesia. The differences between groups 1 and 2, with regard to anesthesia type and arrhythmia, were statistically significant (P <0.001). No patients required additional medical therapy or hospitalization for cardiac events, and all patients safely completed the SWL treatment.
Conclusions:
With careful monitoring and adequate precautions, external fixed-rate signal generators are safe, effective methods for pacing the Medstone shock wave lithotripter and probably other extracorporeal shock wave lithotripters.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
08:43Advanced Cardiac Rhythm Management by Applying Optogenetic Multi-Site Photostimulation in Murine Hearts
Published on: August 26, 2021
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Mechanism of Cardiac Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias III: Characteristics of Dysrhythmias
Dysrhythmias VI: Management of Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias