Frame slippage verification in stereotactic radiosurgery

K Otto1, B G Fallone

  • 1Medical Physics Unit, McGill University, Montreal General Hospital, PQ, Canada.

International Journal of Radiation Oncology, Biology, Physics
|May 20, 1998
PubMed
Summary

This study introduces a 3D image matching method to detect frame slippage in stereotactic radiosurgery (SRS). The technique accurately verifies patient positioning, ensuring treatment precision.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Retraction: Minimal skin dose increase in longitudinal rotating biplanar linac-MR systems: examination of radiation energy and flattening filter design (2016<i>Phys. Med. Biol.</i>61 3527).

Physics in medicine and biology·2025
Same author

Robotic-assisted treatment of obturator nerve entrapment 5 years after retropubic tension-free vaginal tape insertion.

Facts, views & vision in ObGyn·2024
Same author

Transgenerational transmission of psychopathology: when are adaptive emotion regulation strategies protective in children?

Child and adolescent psychiatry and mental health·2024
Same author

Ovarian Dysgerminoma - Challenging Presurgical Diagnosis and Mini-Mally Invasive Treatment.

Archives of clinical and medical case reports·2023
Same author

The CoPhyLab comet-simulation chamber.

The Review of scientific instruments·2021
Same author

Feasibility of energy adaptive angular meshing for perpendicular and parallel magnetic fields in a grid based Boltzmann solver.

Biomedical physics & engineering express·2021

Area of Science:

  • Medical Physics
  • Radiotherapy Technology
  • Image Analysis

Background:

  • Frame slippage in stereotactic radiosurgery (SRS) can compromise treatment accuracy.
  • Accurate patient positioning is critical for effective SRS delivery.
  • Existing methods for detecting frame slippage may have limitations.

Purpose of the Study:

  • To develop and evaluate a novel 3D image matching method for detecting frame slippage in SRS.
  • To improve the accuracy and reliability of patient positioning verification.
  • To utilize Digitally Reconstructed Radiographs (DRRs) interactively matched to portal images.

Main Methods:

  • Superimposing 3D DRRs onto edge-detected portal images obtained before treatment.
  • Interactively manipulating CT data in 3D (rotations, translations) to generate new DRRs.

Related Experiment Videos

  • Matching anatomical structures and using a fiducial marker array for verification.
  • Evaluating the method with simulated portal images and a radiographic head phantom.
  • Main Results:

    • The 3D matching method achieved a mean radial alignment error of 0.82 mm in simulations, significantly better than conventional techniques (3.52 mm).
    • Phantom tests showed mean alignment displacement errors of approximately 1 mm in each stereotactic coordinate (x, y, z).
    • The overall mean radial error in phantom tests was 1.94 mm (SD = 0.61 mm).

    Conclusions:

    • The developed 3D image matching system demonstrates accuracy suitable for stereotactic radiosurgery.
    • This method serves as an effective tool for verifying frame slippage during SRS.
    • The findings support the clinical utility of this technique for enhancing treatment safety and precision.