An evaluation of a modified Macintosh laryngoscope in a manikin

M J Bucx1, C J Snijders, M H van der Vegt

  • 1Department of Anaesthesia, Academic Medical Center, University of Amsterdam, Erasmus University Rotterdam, The Netherlands.

Abstract

Insights

A modified Macintosh laryngoscope blade significantly reduced forces on maxillary incisor teeth during intubation. This innovation maintained high intubation success rates, offering a safer option for dental protection during airway management.

Area of Science:

  • Anesthesiology
  • Medical Device Engineering

Background:

  • Dental trauma is a risk during endotracheal intubation.
  • Macintosh laryngoscope blades are commonly used for airway management.
  • Modifications to laryngoscope blades may reduce forces on teeth.

Purpose of the Study:

  • To compare a modified Macintosh blade with a reduced proximal flange against the original blade.
  • To evaluate the forces exerted on maxillary incisor teeth during laryngoscopy.
  • To assess the impact on tracheal intubation success rates.

Main Methods:

  • A manikin study utilizing sensors to measure axial and transverse forces on maxillary incisors.
  • Fourteen staff anesthetists and 16 residents performed laryngoscopies with both blade types.
  • Each participant performed two laryngoscopies with each blade.

Main Results:

  • All laryngoscopies achieved successful tracheal intubation.
  • The modified blade reduced maximal axial forces by 12 N (P < 0.0005).
  • The modified blade reduced mean axial forces by 5.8 N (P < 0.0005).

Conclusions:

  • The modified Macintosh blade effectively reduces forces on maxillary incisors.
  • Intubation success rates were comparable between the modified and original blades.
  • Manikin studies are valuable for comparative laryngoscope testing.