Multi-step Preparation Technique to Recover Multiple Metabolite Compound Classes for In-depth and Informative Metabolomic Analysis
Published on: July 11, 2014
Simultaneous variation of temperature and gradient steepness for reversed-phase high-performance liquid
J W Dolan1, L R Snyder, D L Saunders
1LC Resources Inc., Walnut Creek, CA 94596, USA.
Abstract:
The preceding paper (Part I) suggests that simply optimizing temperature and gradient steepness will often provide an adequate reversed-phase HPLC separation. In some cases, however, this procedure will prove unsuccessful, and then further method-development experiments (involving change in other separation conditions) will be required. One strategy is to change a variable other than temperature or gradient steepness, followed by re-optimization of the latter two variables. The present paper examines the application of this approach with the aid of computer simulation to several samples.
Related Concept Videos
Chromatography: Introduction
The phase in which the compounds linger or on which the compounds adsorb is called the stationary phase, whereas the mobile phase is the solvent that carries the solutes to be analyzed. In traditional column chromatography, the mixture flows through the stationary phase, and the compounds partition between the stationary and mobile phases...
Chromatographic Methods: Terminology
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Instrumentation