Related Experiment Videos
Flow visualization as a complementary tool to hemolysis testing in the development of centrifugal blood pumps
T Yamane1, B Asztalos, M Nishida
1Mechanical Engineering Laboratory, Tsukuba, Japan.
Abstract:
With a 250% scaled-up pump model, high speed video camera, and argon ion laser light sheet, flow patterns related to hemolysis were visualized and analyzed with 4 frame particle tracking software. Different flow patterns and shear distributions were clarified by flow visualization for pumps modified to have different hemolysis levels. A combination of in vitro hemolysis tests, flow visualization, and CFD analysis suggested a close relationship between hemolysis and high shear caused by small impeller/casing gaps. Because arbitrary cross sections can be illuminated by laser light sheet, flow visualization is a useful tool in finding locations related to hemolysis in the design process of rotary blood pumps.