Contrast sensitivity after implantation of diffractive bifocal and monofocal intraocular lenses
E Haaskjold1, E D Allen, R L Burton
1Eye Department, National Hospital, Oslo, Norway.
Purpose:
To compare contrast sensitivity (CS) after implantation of a diffractive bifocal intraocular lens (IOL) and a monofocal IOL of similar design.
Setting:
Seven European centers.
Methods:
In this randomized, prospective study, CS was tested 5 months after cataract and IOL implantation surgery in 115 patients with a diffractive bifocal IOL and 106 patients with a monofocal IOL. It was also tested in a subgroup of 38 patients who had bilateral implantation of a diffractive bifocal IOL. Contrast sensitivity was tested using the Vision Contrast Test System (VCTS).
Results:
In patients with a best corrected visual acuity (BCVA) of 1.0 or better, the CS at all spatial frequencies (1.5 to 18 cycles/degree), both at distance and near, was slightly lower in the bifocal IOL group than in the monofocal group. Mean values were within the normal range. In patients with a BCVA of less than 1.0, the CS was lower and the difference between the bifocal and monofocal groups was less. In patients with bilateral bifocal IOLs, CS was better when tested bilaterally than when testing the better eye alone. Pupil size affected the results to a small degree. Contrast sensitivity appeared to improve over time after implantation of a diffractive bifocal IOL.
Conclusions:
In patients with cataract and no other eye pathology, the diffractive bifocal IOL with slightly reduce the CS at all spatial frequencies. In those with reduced visual acuity after cataract surgery, CS will be reduced accordingly. In this situation, the reduction from the diffractive bifocal optic would be minor.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...