A Novel Application of Musculoskeletal Ultrasound Imaging
Published on: September 17, 2013
A comparison of ultrasonic and mechanical stadiometry
V Watt1, M Pickering, J K Wales
1Medical School, University of Sheffield.
Aim:
To compare an ultrasonic height measuring device (Gulliver) with mechanical stadiometry and the classical "book and tape measure" method.
Methods:
Blinded duplicate measurements of height were made on each of 14 children by a pair of observers using a stadiometer (H) and Gulliver (G). Height was measured on a further 18 children by parents and an auxologist using Gulliver and the book and tape method (TM), and the results were compared with those obtained with a single stadiometry measurement. Finally, measurement of a rigid metal box was made on 10 occasions by the three methods.
Results:
In the group of 14 children, the mean difference (range) in height (H minus G) was +2.8 cm (+0.5 to +4.55 cm), with H giving a systematically higher value in 276 of 280 individual measurements. In the group of 18 children, height by H was greater than by G or TM in 47 of 52 individual measurements. The mean (SD) height of the box by H (61.60 (0.07) cm) was greater than by G (60.96 (0.15) cm; p < 0.001) but not TM (61.4 (0.16) cm; p > 0.05). G and TM produced three times less reliable estimations of height than H, but with a large difference in cost, and there was evidence of systematic underrecording of height by 0.5 cm with G.
Conclusions:
Stadiometry is precise and reproducible, and can detect true changes in height over one month periods in mid-childhood, and should remain the standard way of observing growth. The book and tape method can produce clinically acceptable quarterly estimations of height that can be performed in the home.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Electronic Distance Measuring Instruments