An automated chromogenic peptide substrate assay for coagulation factor XII
D W Jones1, M J Gallimore, M Winter
1Haemophilia Comprehensive Care Centre, Kent and Canterbury Hospitals NHS Trust, Canterbury, Kent, UK. wynne.jones@kch-tr.sthames.nhs.uk
Summary
An automated assay for factor XII (FXIIcs) accurately measures FXII levels, proving insensitive to lupus anticoagulants. This method enables reliable FXII testing in large patient cohorts, aiding thrombosis and coagulation research.
Area of Science:
- Hematology
- Clinical Chemistry
- Coagulation Diagnostics
Background:
- Factor XII (FXII) plays a role in the intrinsic coagulation pathway.
- Accurate measurement of FXII levels is crucial for diagnosing bleeding and thrombotic disorders.
- Existing FXII assays may be affected by lupus anticoagulants, complicating interpretation.
Purpose of the Study:
- To develop and validate an automated chromogenic peptide substrate assay for factor XII (FXIIcs).
- To compare FXIIcs assay results with conventional clotting (FXIIct) and immunochemical (FXIIag) assays.
- To assess the performance of the FXIIcs assay in blood donors, patients with venous thrombosis, and lupus anticoagulant-positive (LA+) patients.
Main Methods:
- Development of an automated FXIIcs assay on a Cobas Mira S Plus analyzer using a commercial kit.
- Analysis of plasma samples from 320 blood donors, 206 venous thrombosis patients, and 74 LA+ patients.
- Comparative analysis of FXIIcs results against FXIIct and FXIIag assays.
Main Results:
- High correlation (r=0.92) observed between FXIIcs and FXIIct assays in blood donors.
- Low FXII levels detected in a small percentage of blood donors and venous thrombosis patients, all lupus anticoagulant-negative.
- Significant discrepancies between FXIIcs and FXIIct in LA+ patients, with FXIIag results aligning with FXIIcs.
Conclusions:
- The automated FXIIcs assay provides accurate and reliable measurement of factor XII levels.
- The FXIIcs assay is insensitive to lupus anticoagulants, overcoming limitations of FXIIct assays.
- This automated method facilitates routine and accurate FXII level determination in large patient populations.