Long-term follow-up of the antheor inferior vena cava filter

S R Harries1, I P Wells, C A Roobottom

  • 1Imaging Directorate, Derriford Hospital, Plymouth, UK.

Clinical Radiology
|June 18, 1998
PubMed

Insights

The Antheor inferior vena cava (IVC) filter showed significant long-term complications, including migration and fracture. Due to these safety concerns, the IVC filter has been withdrawn from clinical use.

Area of Science:

  • Medical device safety
  • Vascular surgery

Background:

  • Inferior vena cava (IVC) filters are used to prevent pulmonary embolism.
  • The Antheor IVC filter was one such device used clinically.

Purpose of the Study:

  • To present the first long-term follow-up of the Antheor IVC filter.
  • To evaluate the safety and efficacy of the Antheor IVC filter in clinical use.

Main Methods:

  • A retrospective study of 20 patients who received the Antheor IVC filter between 1994 and 1996.
  • Long-term follow-up was conducted on surviving patients.

Main Results:

  • One filter migrated to the pulmonary artery, causing death.
  • Three of 11 assessed filters showed fracture.
  • One filter had associated thrombus, and one IVC was occluded.

Conclusions:

  • The Antheor IVC filter is associated with a high incidence of central migration and filter fracture.
  • These findings led to the withdrawal of the Antheor IVC filter from clinical use.