Measuring Ascending Aortic Stiffness In Vivo in Mice Using Ultrasound
Published on: December 2, 2014
A comparison between systolic aortic root pressure and finger blood pressure
E G Philippe1, J L Hébert, C Coirault
1Service de Physiologie Cardio-Respiratoire, CHU de Bicêtre, Le Kremlin Bicêtre, France.
Background:
Digital photoplethysmography is used to assess hemodynamic variability and baroreflex sensitivity. Numerous studies have critically evaluated the accuracy of the photoplethysmographic device against peripheral pressure. The aim of our study was to compare finger blood and aortic root pressure.
Methods:
We prospectively compared simultaneous recordings of systolic pressure at the aortic root and finger level over three consecutive respiratory cycles in 15 patients (56+/-11 years) undergoing routine cardiac catheterization. Data were obtained at baseline, during deep breathing maneuver (0.1 Hz), and after left ventricular cineangiography.
Results:
At baseline, systolic finger pressure overestimated systolic aortic pressure (145.2+/-22.5 vs 115.0+/-20.1 mm Hg; p<0.001). The pressure difference (30.2+/-17.0 mm Hg) was not influenced by systolic aortic pressure. There was no relationship between pressure difference and the main determinants of the pulse wave amplification phenomenon. There was a beat-to-beat relationship between finger and aortic pressure in 14 of 15 subjects (slope ranging from 0.37 to 1.70; ordinate: from -56 to +98 mm Hg). During the deep breathing maneuver and after left ventricular cineangiography, finger pressure still overestimated aortic pressure by 32.3+/-15.0 mm Hg and 38.3 13.9 mm Hg, respectively (each p<0.001). There was a beat-to-beat relationship between systolic aortic root pressure (IAoBP) and systolic finger (FBP) in 13 of 15 patients, with major scattering of both slopes and ordinates. Throughout the study, there was no predictable relationship between the level of IAoBP and pressure bias.
Conclusions:
As expected, FBP was almost always higher than IAoBP. Importantly, the differences in systolic pressure did not correlate with known determinants of the pulse wave amplification phenomenon. The device must be used cautiously if one wants to noninvasively track spontaneous or induced changes in IAoBP.
Related Concept Videos
Blood Pressure
The average BP in an adult is typically around 120/80 mmHg (millimeters of mercury). In this measurement, the numerator (120) indicates the systolic pressure, which is the pressure in the arteries during the contraction of the heart's ventricles as blood is expelled. The denominator (80) represents the diastolic...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Blood Pressure
Measurement of Blood Pressure