[An experience with St. Jude medical prosthetic 19A-HP which resulted in restricted opening at an early stage after

K Mashiko1, S Ishii, H Naganuma

  • 1Department of Cardiovascular Surgery, Fuji City Hospital, Shizuoka, Japan.

Insights

Aortic valve replacement using the St. Jude Medical Hemodynamic Plus prosthesis showed leaflet restriction. This complication may stem from structural issues in the prosthesis, potentially due to implantation forces.

Area of Science:

  • Cardiovascular Surgery
  • Biomedical Engineering
  • Prosthetic Valve Design

Background:

  • Aortic valve replacement (AVR) is a common procedure for aortic stenosis and regurgitation.
  • The St. Jude Medical Hemodynamic Plus (HP) series is designed to improve orifice area in small aortic roots.
  • This case involves AVR using a 19A-HP prosthesis in a patient with a small aortic orifice (18.9 mm).

Observation:

  • The patient underwent uneventful AVR with a 19A-HP prosthesis.
  • Nine days post-operation, cinefluoroscopy revealed symmetrical restriction of the prosthetic valve leaflets.
  • Doppler echocardiography showed a significant left ventricular-aortic (LV-Ao) pressure gradient of 60 mmHg, despite the absence of symptoms.

Findings:

  • The symmetrical nature of the leaflet restriction suggests a structural issue with the prosthesis rather than surgical technique.
  • The unreinforced orifice ring of the HP prosthesis may be susceptible to deformation from longitudinal forces.
  • It is hypothesized that the forceful insertion of the 19A-HP prosthesis, potentially instead of a smaller size (17-HP), induced these forces, altering the hinge mechanism and restricting leaflet motion.

Implications:

  • This case highlights a potential structural vulnerability in the St. Jude Medical HP series, specifically the orifice ring's susceptibility to deformation.
  • The findings suggest that careful consideration of prosthesis size and implantation technique is crucial to avoid inducing forces that could compromise valve function.
  • Further investigation into the biomechanics of HP prosthesis implantation and long-term structural integrity is warranted to prevent such complications.