Rapid arterial anastomosis with titanium clips
Related Experiment Video
Updated: Oct 29, 2026

Evaluation of a Novel Laser-assisted Coronary Anastomotic Connector - the Trinity Clip - in a Porcine Off-pump Bypass Model
Published on: November 24, 2014
1Department of Surgery, Uniformed Services University of the Health Sciences, Bethesda, Maryland, USA.
Background:
The obvious advantages of rapid arterial anastomoses have prompted the continuing search for more rapid anastomotic techniques to complement the standard sutured anastomosis. Nonpenetrating, arcuate-legged titanium vascular closure staple (VCS) clips were initially developed for microvascular anastomoses. The purpose of this study was to compare VCS clips with sutured arterial end-to-end anastomosis in larger vessels.
Methods:
In 6 pigs, transacted iliac arteries were reanastomosed with VCS clips on one side and continuous 6-0 polypropylene suture on the other.
Results:
The reconstruction time was 8.4+/-5.2 minutes for clip closure and 12.0+/-6.6 minutes for suture closure (P = 0.033). All vessels were patent half an hour after completing the anastomoses with no signs of early thrombosis.
Conclusions:
Arterial end-to-end anastomosis can be performed more rapidly with VCS clips than continuous sutures, and are potentially useful in situations where the clamp time of the vessel is critical.