Determination of ascorbic acid in nanolitre samples by means of capillary batch injection analysis
U Backofen1, F M Matysik, W Hoffmann
1Forschungszentrum Karlsruhe, Institut für Instrumentelle Analytik, Germany.
Biomedical Chromatography : BMC
|July 1, 1998
Abstract:
Capillary batch injection analysis (CBIA) is applied to the determination of ascorbic acid in nanolitre samples at neutral pH. It is demonstrated that pulsed amperometric detection is superior to constant-potential amperometric detection with respect to signal stability. The calibration plot is linear within the studied concentration range between 0.5 mM and 5 mM and the reproducibility is characterized by a relative standard deviation of 0.80% for 10 successive injections of 188 nL of 3.5 mM ascorbic acid. The attainable sample throughput is up to 500 per hour.