A simple algorithm for planar image registration in radiation therapy

J Cai1, J C Chu, V A Saxena

  • 1Department of Medical Physics, Rush Presbyterian St. Luke's Medical Center, Rush University, Chicago, Illinois 60612, USA. cai@therad.rpslmc.edu

Medical Physics
|July 3, 1998
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

The tumor suppressor NOR1 suppresses cell growth, invasiveness, and tumorigenicity in glioma.

Neoplasma·2020
Same author

[Epidemiological characteristics of COVID-19 family clustering in Zhejiang Province].

Zhonghua yu fang yi xue za zhi [Chinese journal of preventive medicine]·2020
Same author

Reply by Authors.

The Journal of urology·2020
Same author

[Karyomegalic interstitial nephritis: report of a case].

Zhonghua bing li xue za zhi = Chinese journal of pathology·2019
Same author

Intraspinal Paragonimiasis in Children: MRI Findings and Suggestions for Pathogenesis.

AJNR. American journal of neuroradiology·2019
Same author

Applications of Ultrafine Limestone Sorbents for the Desulfurization Process in CFB Boilers.

Environmental science & technology·2019
Same journal

Correction to "On the shape of the radiation survival curve in tumor spheroids: The role of oxygen heterogeneity".

Medical physics·2026
Same journal

Multi-view constrained semi-supervised vertebra detection for 3D ultrasound spine volume.

Medical physics·2026
Same journal

Accuracy of quantitative <sup>177</sup>Lu SPECT/CT imaging: A systematic review.

Medical physics·2026
Same journal

Physics-constrained dual-domain network for CBCT reconstruction from orthogonal X-rays in gynecologic radiotherapy.

Medical physics·2026
Same journal

Decomposition-based harmonization for quantitative PET imaging across scanners and radiotracers.

Medical physics·2026
Same journal

Development and evaluation of an in vivo dose-based monitoring system for electron FLASH radiation therapy.

Medical physics·2026
See all related articles

This study introduces a simplified algorithm for planar image registration, enhancing patient setup verification in radiation therapy. The new method achieves high accuracy, with typical errors under 1 mm and 1 degree, comparable to existing techniques.

Area of Science:

  • Medical Physics
  • Radiotherapy
  • Image Analysis

Background:

  • Accurate patient setup is crucial for effective radiation therapy.
  • Image registration plays a key role in verifying patient positioning before treatment delivery.
  • Existing methods for image registration can be complex and computationally intensive.

Purpose of the Study:

  • To develop a simplified and accurate algorithm for planar image registration.
  • To apply the algorithm for simulator and portal image registration in radiation therapy patient setup verification.
  • To evaluate the algorithm's performance against established methods.

Main Methods:

  • The algorithm simplifies curve registration by matching point sets along curves, integrating Balter et al.'s three steps into a single procedure.

Related Experiment Videos

  • It utilizes relative path length shift as a fitting parameter and accounts for different magnification factors.
  • The method avoids complex curvature calculations, making it computationally efficient.
  • Main Results:

    • Numerical simulations and clinical/phantom images demonstrate the algorithm's accuracy.
    • Typical registration errors are less than 1 mm in translation and 1 degree in rotation.
    • Comparison with the chamfer method shows agreement within 0.5 mm and 0.5 degrees.

    Conclusions:

    • The presented algorithm offers a simple yet accurate solution for planar image registration.
    • It is effective for patient setup verification in radiation therapy, improving upon existing methods.
    • The algorithm's efficiency and accuracy make it a valuable tool in clinical practice.