Storage telemetric system for detection of radionuclide tracers in humans
Abstract:
An external-storage telemetric device for continuous detection and registration of radionuclide tracers in humans is described. The weight of the system is 600 g. A pocket-sized ECG cassette tape recorder with 24-h continuous registration constitutes the storage unit. Two end-window GM detectors (Philips 18515) are used as surface detectors. Both gamma- and beta-emitters are detected by this equipment. The linear range is 100 counts/s. Renal excretion of [125I]iodide during 30 h in a normal person is shown. The disappearance rates are 0.1 h-1 during the day and 0.05 h-1 during the night. The coefficient of variation of the individual slopes has been calculated to less than 5%. A comparison with the analog radiotelemetric equipment is shown. The storage telemetric equipment can be applied outside the hospital, when the patient is ambulatory.
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET

![Automated Radiochemical Synthesis of [18F]3F4AP: A Novel PET Tracer for Imaging Demyelinating Diseases](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F55537.jpg&w=3840&q=50)