Related Experiment Videos
Determination of polyamines in human prostate by high-performance liquid chromatography with fluorescence detection
1Central Laboratory, Affiliated Hospital of Shandong Medical University, Jinan, China.
Abstract:
A high-performance liquid chromatographic method for the determination of polyamines in human prostate has been developed. This method is based on pre-column derivatization with dansyl chloride (Dns-Cl). The derivatives were separated on a muBondapak C18 column (250x4.6 mm I.D.; 10 microm), and eluted with methanol and distilled water using a one-step linear gradient. The column eluate was monitored by fluorescence detection (excitation, 370 nm; emission, 506 nm). The within-assay precision of the study (C.V.) was as follows: putrescine (PUT) 2.88%, spermidine (SPD) 2.94% and spermine (SP) 1.17%. The between-assay precision (C.V.) was: PUT 2.66%, SPD 3.06%, SP 2.79%. The recovery was greater than 97%. The detection limit for PUT, SPD and SP were 0.05, 0.08 and 0.06 nmol/ml, respectively. In contrast to other studies, sample or polyamine derivatives did not require extraction with an organic solvent such as ethanol, evaporation under vacuum or other condensation procedures. This is a simple, rapid and sensitive method that can be applied to the determination of polyamines in nearly all biological tissues and body fluids, such as urine and serum.