Evaluation of a new low-dose digital x-ray device: first dosimetric and clinical results in children
G Kalifa1, Y Charpak, C Maccia
1Service de Radiologie, Hôpital Saint-Vincent de Paul, F-75 674 Paris Cedex 14, France.
Background:
A new low-dose digital X-ray device, based on Charpak's Nobel prize-winning multiwire chamber, enables the production of images at very low doses. Objectives. To present the first dosimetric and clinical results.
Materials And Methods:
The analysis was performed on 93 children with scoliosis and 47 undergoing pelvic radiography. The comparative study between conventional X-ray and the new technique focused on three points: (1) the dose delivered by each system (2) the diagnostic information provided by each system and (3) comparison of image quality criteria with European guidelines.
Results:
The mean ratio of conventional dose to that of the low-dose technique was 13.1 for the spinal examination and 18.8 for the pelvis. There was no significant difference in diagnostic information available from each modality, but there was a slight difference in quality criteria in favour of the conventional technique.
Conclusion:
This new device allows spectacular dose reduction, consistent with adequate clinical information. Improvements of the prototype will lead to extension of potential indications and industrial development.