A portable system for closed loop control of the paralysed hand using functional electrical stimulation
Related Experiment Video
Updated: Oct 29, 2026

Use of a Foot-Induced Digitally Controlled Resistance Device for Functional Magnetic Resonance Imaging Evaluation in Patients with Foot Paresis
Published on: July 7, 2023
1Department of Medical Physics and Biomedical Engineering, Salisbury District Hospital, UK.
Abstract:
A portable and closed-loop system is described for the paralysed hand using transcutaneous electrical stimulation. It is implemented using a modest microprocessor which receives data from force sensors mounted in a glove on the users hand. A display shows parameter values and a menu for the user to sequentially select controller states. For the grip state, the control loop is basically proportional plus a two stage integral response (gain adaptation). Eight channels can be accommodated in the stimulator. The system was evaluated with the help of a tetraplegic who managed to hold everyday objects in a stable grip.