A portable system for closed loop control of the paralysed hand using functional electrical stimulation
1Department of Medical Physics and Biomedical Engineering, Salisbury District Hospital, UK.
Medical Engineering & Physics
|July 17, 1998
Summary
This study presents a portable, closed-loop system using transcutaneous electrical stimulation to restore hand grasp for paralyzed individuals. A tetraplegic user successfully held objects, demonstrating the system's effectiveness in improving hand function.
Area of Science:
- Neuroprosthetics
- Rehabilitation Engineering
- Biomedical Devices
Background:
- Paralysis, particularly in the hand, significantly impacts daily living and independence.
- Existing assistive technologies for hand paralysis often lack portability or intuitive control.
- Restoring functional grasp is a key goal in neurorehabilitation.
Observation:
- A portable, closed-loop system was developed for hand grasp restoration.
- The system utilizes transcutaneous electrical stimulation controlled by a microprocessor and force sensors in a glove.
- A user interface allows for sequential selection of controller states, including a proportional plus two-stage integral response for grip control.
Findings:
- The system successfully enabled a tetraplegic individual to achieve a stable grip on everyday objects.
- The closed-loop control, incorporating gain adaptation, facilitated consistent grasp force.
- The portable design and user-friendly interface demonstrated practical applicability.
Implications:
- This technology offers a promising avenue for restoring hand function and independence in individuals with paralysis.
- The portable and closed-loop nature of the system suggests potential for widespread clinical and home use.
- Further development could enhance control capabilities and expand applications for upper limb neuroprosthetics.