Related Experiment Video
Updated: Aug 15, 2026

Sequential Application of Glass Coverslips to Assess the Compressive Stiffness of the Mouse Lens: Strain and Morphometric Analyses
Published on: May 3, 2016
Measurement of axial length in the calculation of intraocular lens power
Purpose:
Accurate measurement of ocular axial length is essential for accurate intraocular lens (IOL) power calculation. Although it is common practice to average several axial length measurements to improve accuracy, it has been suggested that a single high-quality A-scan ultrasonographic measurement is adequately accurate owing to the high test-retest reliability of A-scan biometry. The aim of this study was to compare the accuracy of a single high-quality A-scan measurement with that of the average of three acceptable measurements in the calculation of IOL power.
Method:
We studied 103 eyes of 103 patients who underwent cataract-IOL surgery. All these patients underwent pre-operative ocular biometry, a standardised extracapsular cataract extraction with posterior chamber IOL implantation, and clinical refraction between 10 and 14 weeks post-operatively.
Results:
There was no statistically significant difference between the two study groups in measured axial length (p > 0.01), calculated emmetropic IOL power (p > 0.05) or the prediction of post-operative refraction (p > 0.99).
Conclusion:
The use of a single high-quality axial length measurement was as accurate as the mean of three acceptable axial length measurements in the calculation of IOL power.
More Related Videos
06:55Scanning Light Scattering Profiler (SLPS) Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
07:44In vivo Structural Assessments of Ocular Disease in Rodent Models using Optical Coherence Tomography
Published on: July 24, 2020
Related Concept Videos
Focusing of Light in the Eye
Eccentric Axial Loading in a Plane of Symmetry