Microscopic evaluation of three endodontic files pre- and postinstrumentation
I Bonetti Filho1, R Miranda Esberard, R de Toledo Leonardo
1Computing Resources Department, University of Texas Health Science Center at San Antonio 78284-7892, USA.
Abstract:
K-type files, Sureflex NiTi files, and FlexoFiles were evaluated morphometrically as they come from the manufacturer, and after being used to instrument root canals 1, 3, and 5 times in maxillary premolars. Evaluation was made using a stereomicroscope at x40 magnification and photomicrographs. The abnormalities found after instrumentation were submitted for statistical analysis. It was concluded that small-sized stainless-steel instruments should be discarded after one use. On the other hand, #30 stainless-steel K-files could be used up to three times and #30 stainless-steel FlexoFiles up to five times. Sureflex NiTi instruments, even after five times, did not show appreciable abnormalities in shape. It is strongly recommended that small-sized nickel-titanium instruments should be discarded after five uses.